The QardioArm Blood Pressure App for Self-Measurement in an Obese Population: Validation Study Using the European Society of Hypertension International Protocol Revision 2010

dc.contributor.authorMazoteras Pardo, Victoria
dc.contributor.authorBecerro de Bengoa Vallejo, Ricardo
dc.contributor.authorLosa Iglesias, Marta Elena
dc.contributor.authorLópez López, Daniel
dc.contributor.authorPalomo López, Patricia
dc.contributor.authorRodríguez Sanz, David
dc.contributor.authorCalvo Lobo, César 
dc.date.accessioned2020-02-20T19:20:34Z
dc.date.available2020-02-20T19:20:34Z
dc.date.issued2018
dc.description.abstractBackground: Obesity and high blood pressure (HBP) pose high cardiovascular risks, and they are frequent causes of cardiovascular disease. Objective: The aim of this study was to validate the mobile app QardioArm for high blood pressure monitoring in obese subjects (body mass index ≥30 kg/m2) according to guidelines in the European Society of Hypertension-International Protocol 2 (ESH-IP2). Methods: We recruited 33 obese subjects and measured their blood pressure using QardioArm (test device) and Omron M3 Intellisense (Omron Healthcare, Kyoto, Japan; standard device). We compared systolic blood pressure (SBP), diastolic blood pressure (DBP), and heart rate (HR) according to the ESH-IP2. Results: A total of 95 of 99 differences for SBP and 91 of 99 for DBP displayed absolute differences within 10 mm Hg. A total of 98 of 99 differences for SBP and 98 of 99 for DBP exhibited absolute differences within 15 mm Hg. This result satisfied requirements for part 1 of the ESH-IP2. A total of 27 out of 33 individuals for SBP and 30 out of 33 individuals for DBP had a minimum of 2 of 3 comparisons within 5 mm Hg difference. None of the subjects had 3 differences outside 5 mm Hg for SBP and DBP, satisfying part 2 of the ESH-IP2. For HR measurements, a total of 90 of 99 differences had absolute differences within 3 beats per minute (bpm), and a total of 94 or 99 differences had absolute differences within 5 bpm. A total of 98 of 99 differences had absolute differences within 8 bpm. Therefore, the test device satisfied part 1 of ESH-IP2 criteria for HR. For part 2 of ESH-IP2, 31 of 33 individuals had a minimum of 2 of 3 comparisons within 3 bpm difference for HR. Only 1 of 33 subjects had 3 differences outside 3 bpm. Conclusions: To the best of our knowledge, this was the first study to show that an app that measures blood pressure and HR meets the requirements of the ESH-IP2 in an obese population. We believe the ESH-IP2 should publish explicit criteria for validation of blood pressure devices in specific populations.spa
dc.description.filiationUEMspa
dc.description.impact4.301 JCR (2018) Q1, 2/26 Medical Informatics, 13/98 Health Care Sciences & Servicesspa
dc.description.impactNo data SJR 2018spa
dc.description.impactNo data IDR 2018spa
dc.description.sponsorshipSin financiaciónspa
dc.identifier.citationMazoteras-Pardo, V., Becerro-De-Bengoa-Vallejo, R., Losa-Iglesias, M. E., López-López, D., Palomo-López, P., Rodríguez-Sanz, D., & Calvo-Lobo, C. (2018). The QardioArm Blood Pressure App for Self-Measurement in an Obese Population: Validation Study Using the European Society of Hypertension International Protocol Revision 2010. JMIR MHealth and UHealth, 6(10), e11632. https://doi.org/10.2196/11632spa
dc.identifier.doi10.2196/11632
dc.identifier.issn2291-5222
dc.identifier.urihttp://hdl.handle.net/11268/8635
dc.language.isoengspa
dc.peerreviewedSispa
dc.rightsAttribution-NonCommercial-NoDerivatives 4.0 Internacional*
dc.rights.accessRightsrestricted accessspa
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/4.0/*
dc.subject.uemObesidadspa
dc.subject.uemPresión sanguíneaspa
dc.subject.unescoObesidadspa
dc.subject.unescoSistema cardiovascularspa
dc.titleThe QardioArm Blood Pressure App for Self-Measurement in an Obese Population: Validation Study Using the European Society of Hypertension International Protocol Revision 2010spa
dc.typejournal articlespa
dspace.entity.typePublication
relation.isAuthorOfPublication43641780-6ebb-488f-8857-532d1133ace6
relation.isAuthorOfPublication224f44e5-15ae-48f2-8e32-ac0879c24e79
relation.isAuthorOfPublication.latestForDiscovery43641780-6ebb-488f-8857-532d1133ace6

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Rodríguez_Sanz_JMIR_2018.pdf
Size:
177.25 KB
Format:
Adobe Portable Document Format
Description:
Versión del editor