Adjustable Transobturator Male System (ATOMS) Infection: Causative Organisms and Clinical Profile
Loading...
Identifiers
Publication date
Authors
Advisors
Editors
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
Objective
To evaluate the clinical profile and the organisms producing adjustable transobturator male system (ATOMS) infection in a contemporary series.
Methods
Multicenter retrospective study evaluating patients undergoing ATOMS explant for clinical signs of infection from a series of 902 patients treated in 9 academic institutions. Clinical and microbiological data were evaluated.
Results
Infection presented in 24 patients (2.7%). The median age was 73 ± 7yrs and the median interval from ATOMS implantation to explant 11 ± 26.5mo. Infection was diagnosed within 3-months after surgery in 7(29.2%). Scrotal port erosion was present in 6 cases (25%) and systemic symptoms of parenchymatous testicular infection in 2(8.3%). The culture of the periprosthetic fluid was positive in 20(83.3%): 12(50%) Gram-negative bacteria, 9(37.5%) Gram-positive cocci and 1(4.2%) yeast. The most frequent isolates were Enterococcus and Proteus sp. (16.7% each), followed by Pseudomona sp. and S. epidermidis (12.5% each). Methicillin resistant S. aureus was detected only in 1 case (4.2%). Despite the infection 17 patients (70.8%) were satisfied with the implant and 18(75%) received a second device (11 repeated ATOMS and 7 AUS) at a median 9.7 ± 12.6mo after explant. Limitations include retrospective design and lack of microbiological cultures in ATOMS explanted for non-infective cause.
Conclusion
Infection of a prosthetic device is a disturbing complication. A proportion of patients with ATOMS infection is associated to scrotal port erosion and/or parenchymatous urinary tract infection. Enterococcus and Proteus sp. are the most common organisms producing ATOMS infection and this could have implications for the selection of the most appropriate surgical prophylaxis.
Description
UNESCO Subjects
Keywords
Bibliographic reference
Angulo, J. C., Rourke, K., Queissert, F., Giammò, A., Schönburg, S., Cruz, F., Martins, F. E., Gonsior, A., & González-Enguita, C. (2021). Adjustable transobturator male system (Atoms) infection: Causative organisms and clinical profile. Urology, 157, 120-127. https://doi.org/10.1016/j.urology.2021.05.111