Resumen:
This systematic review and meta -analysis (MA) aimed to evaluate the diagnostic validity of portable electromyography (EMG) diagnostic devices compared to the reference standard method polysomnography (PSG) in assessing sleep bruxism. This systematic review was completed in accordance with the Preferred Reporting Items for Systematic Reviews and Meta -Analysis (PRISMA) statement and was registered with PROSPERO prior to the accomplishment of the main search. Ten clinical studies on humans, assessing the diagnostic accuracy of portable instrumental approaches with respect to PSG, were included in the review. Methodological shortcomings were identified by QUADAS-2 quality assessment. The certainty of the evidence analysis was established by different levels of evidence according to the Grading of Recommendations, Assessment, Development, and Evaluation (GRADE) framework. A meta -analysis of diagnostic test accuracy was performed with multiple thresholds per study applying a two-stage random effects model, using the thresholds offered by the studies and based on the number of EMG bruxism events per hour presented by the participants. Five studies were included. The MA indicated that portable EMG dia...