Resumen:
The QuantiFERON-TB Gold Plus (QFT-Plus) assay, which features two antigen-stimulated tubes (TB1 and TB2) instead of a single tube used in previous-generation interferon-gamma release assays (IGRAs), was launched in 2016. Despite this, data regarding the assay's performance in the paediatric setting remain scarce. This study aimed to determine the performance of QFT-Plus in a large cohort of children and adolescents at risk of tuberculosis (TB) in a low-burden setting.
Methods: Cross-sectional, multicentre study at healthcare institutions participating in the Spanish Paediatric TB Research Network, including patients <18 years who had a QFT-Plus performed between September 2016 and June 2020.
Results: Of 1726 patients (52.8% male, median age: 8.4 years), 260 (15.1%) underwent testing during contact tracing, 288 (16.7%) on clinical/radiological suspicion of tuberculosis disease (TBD), 649 (37.6%) during new-entrant migrant screening and 529 (30.6%) prior to initiation of immunosuppressive treatment. Overall, the sensitivity of QFT-Plus for TBD (n=189) and for latent tuberculosis infection (LTBI, n=195) was 83.6% and 68.2%, respectively. The agreement between QFT-Plus TB1 and TB2 antigen tubes...