Resumen:
OBJECTIVES: Feasibility study to evaluate efficacy and safety of Adjustable Transobturator Male System (ATOMS) for male stress urinary incontinence (SUI) after transurethral resection of the prostate.
MATERIALS AND METHODS: Twenty patients were implanted ATOMS for SUI caused by transurethral resection of the prostate with or without radiotherapy. Incontinence severity was evaluated as mild (2 pads/day), moderate (3-5 pads/day) or severe (≥6pads/day), and dryness as none or one security pad/day. Changes in pad-test and pad-count after adjustment were investigated, together with operative parameters, patient satisfaction with the procedure, and number and grade of complications (Clavien-Dindo).
RESULTS: Median age was 76.5years. Five patients received previous pelvic radiation (3 prostate, 2 rectal cancer) and 2 (10%) previous failed artificial urinary sphincter with urethral erosion and mechanical failure, respectively. Preoperative SUI was mild in 4 (20%), moderate in 7 (35%) and severe in 9 (45%). Median filling of the system was 13.5ml. Median pad-test decreased from 375±855ml baseline to 10±31.5ml and pad-count from 4±3 to 0±1.5 after adjustment (1±3fillings). Postoperative SUI distribut...