Resumen:
Aim: To evaluate treatment options after surgical revision of adjustable transobturator male system (ATOMS) and the results of further incontinence implantation.
Materials and methods: A retrospective multicenter study evaluating patients with surgical revision of ATOMS in academic institutions. Causes and factors affecting revision-free interval were studied and also the frequency of device explant and placement of second ATOMS or artificial urinary sphincter (AUS) at surgeon discretion. Operative results, complications (Clavien-Dindo), and efficacy (postoperative pad-test, pad-count, patient satisfaction, and patient global impression of improvement [PGI-I scale]) of each treatment option were compared.
Results: Seventy-eight out of 902 patients (8.65%) with ATOMS underwent surgical revision at 4.1 ± 2.4 years mean follow-up and 75 (8.3%) were explanted. The main causes for revision included persistence of incontinence (35.9%) and scrotal port erosion (34.6%). Independent risk factors of the shortened revision-free interval were previous anti-incontinence surgery (HR, 1.83; 95% CI, 1.06-3.16; p = 0.007) and port erosion (HR, 1.83; 95% CI, 1.06-3.16; p = 0.0027). Fifty-eight (6.4%) receive...